LABORATORY OVEN – Lab. Precision Oven – 400◦C (High Temperature)

  • Construction -: Tripled walled construction.
  •  Insulation -: 4 Inch High grade glass wool insulation ensures stable temperature & reduced energy
    Consumption.
  •  Door -: Single door fitted on heavy cast & chromium plated hinges with spring and roller type latch with spring and roller type latch with asbestos gasket to avoid heat loss.
  •  Air Recirculation -: Motorized Blower with V- BELT drive on top side of the chamber develops unique air flow System which ensures maximum uniform temperature distribution inside the chamber.
    Unique air flow assures quick recovery after door Opening’s. V-Belt drive ensures
    Continuous run of motor on high temperature to avoid damage of motor.
  •  Blower Motor -: 1/4 HP, TEFC, F-class insulation, Single Phase, 1440 RPM, 230 Volts.
  •  Heating System -: Specially designed U shape S.S. Tubular Heating elements used on side walls to
    Maintain Uniform condition inside the chamber. Heaters terminals taken out for wiring
    To avoid burning of heater terminals on high temperature.
  •  Safety -: Limit switch for interlocking connection between Motor & heater.
  •  Supply -: 230 V AC, Single Phase , 50 Hz
  •  Temperature Range -: 50° C to 400°C
  •  Temperature Accuracy -: ±1° C.
  •  Temperature Uniformity -: ± 1 % of set temperature.
  •  Trays -: Specially designed Stainless steel / G.I. wire mesh trays which ensure uniform temperature
    Distribution.
  •  Standard Model -: Inside S.S.304 mirror finish & Outer Mild steel (CRCA) sheet with powder coated.
    OR Inside Mild steel (CRCA) with heat resistant aluminium paint & Outer Mild steel
    (CRCA) Sheet with powder coated.
  •  Control System -: Imported Microprocessor based auto tuned PID controller with CE mark & dual display of Set value& process value for precise control of temperature.

TECHNICAL SPECIFICATION

REQUIREMENT FOR INSTALLATION

  1. Place the chamber where ventilation is good.
  2. The chamber must be operated in an ambient of 22◦ C ~ 30◦ C.
  3. Leave minimum 50 mm. Space from the sides and leave 300 mm. Space from the backside.
  4. Place the chamber horizontally levelled.
  5. Give stabilized electric supply (230V ±5%, 50Hz, 1 Phase, 32 Amp). It is recommended to use a voltage stabilizer (KVA).
  6. Proper grounding of the chamber to get satisfactory operation of chamber.
  7. Do not test any product in the chamber, which is capable of generating combustible mixtures.